Measurement Uncertainty Budget Calculator (GUM)

Enter repeat readings, the display resolution and a table of Type B contributions with distributions, sensitivity coefficients and degrees of freedom to get combined and expanded uncertainty.

Runs in your browser. Your files never leave your device. See how to check

…

How to do it

  1. 1
    Paste repeat readings

    For Type A.

  2. 2
    Enter resolution

    Display resolution of the instrument.

  3. 3
    Fill the budget table

    Value, distribution, divisor, sensitivity, dof.

  4. 4
    Choose the coverage factor

    From t or fixed k = 2.

  5. 5
    Read the result

    u_c, ν_eff, k and U.

About this tool

Every measurement result needs a statement of uncertainty, and the Guide to the Expression of Uncertainty in Measurement (GUM) is the international method for producing it. Calibration labs accredited to ISO/IEC 17025 present it as an uncertainty budget: a table of contributions, each converted to a standard uncertainty, combined and expanded with a coverage factor. This calculator builds such a budget, including Type A repeatability from your readings, the display resolution, the Welch-Satterthwaite effective degrees of freedom and the coverage factor.

Who this is for

Calibration and testing laboratory staff, quality engineers, metrology students and anyone preparing an ISO/IEC 17025 uncertainty statement.

Type A evaluation

From n repeat readings, the standard uncertainty of the mean is s/√n with n − 1 degrees of freedom. Paste your readings and it is added to the budget automatically.

Type B evaluation

Other sources come from certificates, specifications, handbooks or judgement. Convert each to a standard uncertainty: - **Normal**: expanded uncertainty from a certificate divided by its coverage factor (often k = 2). - **Rectangular**: half-width a divided by √3, for limits with equal probability, such as resolution or manufacturer specifications. - **Triangular**: a/√6, when values near the centre are more likely. - **U-shaped**: a/√2, for sinusoidal effects such as temperature cycling.

Resolution

A digital display with resolution d contributes d/(2√3). Enter the resolution and it becomes a rectangular term.

Sensitivity coefficients

Each input affects the result through a sensitivity coefficient c_i (a partial derivative or a unit conversion). The contribution is c_i u_i.

Combining and expanding

u_c = √Σ (c_i u_i)². The Welch-Satterthwaite formula gives effective degrees of freedom ν_eff = u_c⁴ / Σ((c_i u_i)⁴/ν_i), and the coverage factor k comes from Student's t for about 95 % (95.45 %) coverage; with large ν_eff, k = 2. U = k u_c.

Reading the budget

The "share" column shows each source's contribution to u_c². Reducing the largest contributors gives the most improvement.

Good practice

Avoid double counting (for example repeatability already included in a certificate value). Report the result as y ± U with k and the coverage probability.

A simple example of sensitivity

If a length L is measured at temperature t and corrected to 20 °C with expansion coefficient α, the sensitivity of the result to temperature is c = L × α. For a 100 mm steel part (α = 11.5 × 10⁻⁶ /K), c = 0.00115 mm/K.

Worked example

Six readings averaging 100.012 have a Type A uncertainty of 0.00058 with 5 degrees of freedom. With a reference standard of ±0.004 (k = 2, 50 dof), rectangular temperature and drift terms of ±0.003 and ±0.002, and 0.001 resolution, u_c = 0.00296, ν_eff ≈ 224, k = 2.01 and U = ±0.0060.

Limits

Linear models with uncorrelated inputs; correlated or strongly non-linear cases need GUM Supplement methods. Not an official accreditation document.

Worked example

Six readings plus three Type B terms give U = ±0.0060 at k = 2.01.

Quick facts

Frequently asked questions

What is the GUM?

The international guide for evaluating and expressing measurement uncertainty.

What is the difference between Type A and Type B?

Type A uses statistics of repeated readings; Type B uses other information.

Why divide by √3?

That converts a rectangular half-width into a standard uncertainty.

When is k not 2?

When effective degrees of freedom are small; k comes from Student's t.

How should I report the result?

As y ± U with the coverage factor and coverage probability.