Hardness Test Calculator (Brinell, Vickers, Knoop)
Enter the test force and the measured indentation diameter, diagonals or length to get HBW, HV or HK, check the test conditions and estimate the tensile strength of steel.
Runs in your browser. Your files never leave your device. See how to check
…
How to do it
- 1Choose the test
Brinell, Vickers or Knoop.
- 2Enter the force
In newtons.
- 3Enter the indenter size
Ball diameter for Brinell.
- 4Enter the indentation
Diameter, diagonals or length.
- 5Read the hardness
With checks and strength.
About this tool
Indentation hardness tests press a hard indenter into the surface with a known force and measure the size of the mark. Brinell uses a tungsten carbide ball and measures the indentation diameter, giving an average hardness over a large area, suited to castings and forgings. Vickers uses a square diamond pyramid and measures both diagonals, covering everything from soft metals to hardened steels and coatings. Knoop uses an elongated diamond for thin layers and brittle materials. This calculator converts your measurements into HBW, HV or HK with the formulas of the ISO standards, checks the Brinell load ratio and d/D ratio, gives the indentation depth for Vickers and Knoop, and estimates the tensile strength of steels from Brinell hardness.
Who this is for
Materials and quality engineers, heat-treaters, metallurgy students, laboratory technicians and inspectors.
Brinell (ISO 6506-1)
HBW = 0.102 × 2F / (π·D·(D − √(D² − d²))), with F in N and D, d in mm. The force-diameter ratio 0.102·F/D² is chosen by material: 30 for steel and cast iron above 140 HB, 10 for copper alloys and softer cast iron, 5 or 2.5 for aluminium and soft metals. The indentation diameter should be 0.24-0.6 D. A test with a 10 mm ball and 29.42 kN (3,000 kgf) is written HBW 10/3000.
Vickers (ISO 6507-1)
HV = 0.1891 × F / d² (F in N, d the mean diagonal in mm), equivalent to 1.8544 × F(kgf)/d². The indentation depth is about d/7. Loads from 0.098 N (HV 0.01, microhardness) to 980 N (HV 100) are used.
Knoop (ISO 4545-1)
HK = 1.451 × F / L² with L the long diagonal in mm. The shallow indent (depth about L/30) suits thin coatings, case-hardened layers and ceramics.
Estimating tensile strength
For carbon and low-alloy steels, Rm ≈ 3.45 × HB MPa is a useful rough guide (ASTM A370 and SAE J417 give tables). It is not valid for cast iron, non-ferrous metals or cold-worked material.
Worked example
A 10 mm ball with 3,000 kgf leaves a 4.00 mm indentation: HBW 229, d/D = 0.40 (valid), and an estimated tensile strength of about 790 MPa. A Vickers test at 10 kgf with diagonals of 0.43 and 0.45 mm gives about 96 HV.
Converting between scales
Conversions between Brinell, Vickers and Rockwell (ISO 18265, ASTM E140) are empirical and material-specific; use the published tables, not a single formula.
Example uses
A heat-treater checks a batch of shafts after quenching and tempering.
Limits
Measurement quality governs accuracy: polish the surface, measure carefully and respect minimum thickness and spacing. This is not professional engineering advice.
Worked example
HBW 10/3000 with a 4.00 mm indentation: 229 HBW, about 790 MPa tensile.
Quick facts
- Category: Engineering & Construction
- Works in your browser, no sign-up, no upload.
- Your input stays on your device.
- Available in English.
- Content last reviewed:
- Official source: Indentation hardness: Brinell HBW = 0.102·2F/(πD(D − √(D² − d²))) (ISO 6506-1), Vickers HV = 0.1891·F/d² (ISO 6507-1), Knoop HK = 1.451·F/L² (ISO 4545-1), F in N; steel tensile strength ≈ 3.45·HB (ASTM A370)
Frequently asked questions
How is Brinell hardness calculated?
Force divided by the curved surface area of the indentation, in kgf/mm².
What load should I use for steel?
A force-diameter ratio of 30: 3,000 kgf with a 10 mm ball.
How is Vickers hardness calculated?
HV = 1.8544 × F (kgf) / d², with d the mean diagonal in mm.
When is Knoop used?
For thin layers, coatings and brittle materials.
Can I convert HB to tensile strength?
Roughly, for steels: Rm ≈ 3.45 × HB MPa.
Is my data stored?
Test values you enter stay on this device.