How to Do a Bland-Altman Analysis and Report Agreement
Why correlation cannot show agreement, how to get bias and limits of agreement with confidence intervals, and how to read and report a Bland-Altman plot.

Method comparison studies ask a simple question: can a new way of measuring something replace the old one? A cheaper meter, a wearable sensor, a second observer or a quicker lab test is only useful if it gives readings close enough to the established method. The standard way to answer that question is the Bland-Altman analysis, and this guide shows how to run one with the free Bland-Altman Agreement Plot Calculator.
Why correlation is not enough
It is tempting to plot one method against the other and quote a correlation coefficient. The trouble is that correlation measures association, not agreement. If one method always reads 20 units higher than the other, the correlation is perfect, yet the two never give the same answer. Correlation also rises when the range of subjects is wide, regardless of how well the methods agree for any one person.
Martin Bland and Douglas Altman made this point in a famous 1986 paper in The Lancet. Their example compared two peak flow meters in 17 people. The correlation was 0.94, which looks excellent, but individual readings differed by up to 80 litres per minute.
The Bland-Altman approach
For each subject, compute the difference between the two readings and their mean. Plot the difference on the vertical axis against the mean on the horizontal axis. Then add three lines: the mean difference, called the bias, and the bias plus and minus 1.96 standard deviations of the differences. Those outer lines are the 95% limits of agreement. For a new subject, about 95% of differences between the methods are expected to lie between them.
In the peak flow data the bias is −2.1 litres per minute, so the two meters agree on average. But the standard deviation of the differences is 38.8, so the limits of agreement run from about −78 to 74. For an individual patient, one meter could read almost 80 litres per minute above or below the other.
Confidence intervals matter
The bias and the limits are estimates from a sample, so they come with uncertainty. The calculator gives a t-based confidence interval for the bias and a confidence interval for each limit. With 17 subjects the lower limit could plausibly be anywhere from about −113 to −43. Small studies cannot pin the limits down, which is a good reason to plan for at least 50 to 100 pairs when the result will guide practice.
Two formulas exist for the standard error of the limits. The 1999 paper by the same authors gives a slightly more accurate form; the 1986 paper used a simpler one that some software still uses. The tool lets you pick either so you can match a published result.
Decide what is acceptable before you look
The limits of agreement describe how far apart the methods can be. Whether that is acceptable is not a statistical question. Before collecting data, decide on the largest difference that would not change a decision, for example 10 mmHg for a blood pressure monitor in a particular setting. Enter it in the acceptable difference box and the calculator shows whether the limits, and the outer ends of their confidence intervals, fall inside it.
Check the assumptions on the plot
The basic method assumes that the differences are roughly normal and that their size does not change across the range of measurement. Look at the plot. If the points fan out as the mean increases, the disagreement is proportional to the size of the reading. The calculator fits a regression line of difference on mean and reports its slope and p-value as a check.
When the spread grows with the level, switch to percentage differences or log ratios. Log ratios are especially neat because their limits can be back-transformed into statements like "the new method reads between 0.92 and 1.08 times the reference".
Also look for outliers. A single subject far from the rest can widen the limits a lot. Do not delete such points silently; investigate them and report what you did.
Reporting the result
A clear report states the number of pairs, the bias with its confidence interval, the limits of agreement with their confidence intervals, and the acceptable difference chosen in advance. The calculator builds a result sentence in that form that you can copy, and it exports the plot as SVG or PNG and the differences as CSV. Mention whether you used absolute, percentage or log-ratio differences and which order of subtraction.
Common mistakes
The most frequent mistakes are relying on correlation or a paired t-test alone, plotting the difference against one method instead of the mean, ignoring a trend in the differences, and treating repeated measurements on the same person as independent pairs. The last needs a modified analysis that this tool does not attempt.
Finally, remember that agreement studies describe instruments and procedures. They are not medical advice, and decisions about patient care should rest on validated devices and local guidance.