Measurement Uncertainty Budget Calculator (GUM)
Enter repeat readings, the display resolution and a table of Type B contributions with distributions, sensitivity coefficients and degrees of freedom to get combined and expanded uncertainty.
S'executa al teu navegador. Els fitxers mai no surten del dispositiu. Mira com comprovar-ho
…
Com es fa
- 1Paste repeat readings
For Type A.
- 2Enter resolution
Display resolution of the instrument.
- 3Fill the budget table
Value, distribution, divisor, sensitivity, dof.
- 4Choose the coverage factor
From t or fixed k = 2.
- 5Read the result
u_c, ν_eff, k and U.
Sobre aquesta eina
Every measurement result needs a statement of uncertainty, and the Guide to the Expression of Uncertainty in Measurement (GUM) is the international method for producing it. Calibration labs accredited to ISO/IEC 17025 present it as an uncertainty budget: a table of contributions, each converted to a standard uncertainty, combined and expanded with a coverage factor. This calculator builds such a budget, including Type A repeatability from your readings, the display resolution, the Welch-Satterthwaite effective degrees of freedom and the coverage factor.
Who this is for
Calibration and testing laboratory staff, quality engineers, metrology students and anyone preparing an ISO/IEC 17025 uncertainty statement.
Type A evaluation
From n repeat readings, the standard uncertainty of the mean is s/√n with n − 1 degrees of freedom. Paste your readings and it is added to the budget automatically.
Type B evaluation
Other sources come from certificates, specifications, handbooks or judgement. Convert each to a standard uncertainty: - **Normal**: expanded uncertainty from a certificate divided by its coverage factor (often k = 2). - **Rectangular**: half-width a divided by √3, for limits with equal probability, such as resolution or manufacturer specifications. - **Triangular**: a/√6, when values near the centre are more likely. - **U-shaped**: a/√2, for sinusoidal effects such as temperature cycling.
Resolution
A digital display with resolution d contributes d/(2√3). Enter the resolution and it becomes a rectangular term.
Sensitivity coefficients
Each input affects the result through a sensitivity coefficient c_i (a partial derivative or a unit conversion). The contribution is c_i u_i.
Combining and expanding
u_c = √Σ (c_i u_i)². The Welch-Satterthwaite formula gives effective degrees of freedom ν_eff = u_c⁴ / Σ((c_i u_i)⁴/ν_i), and the coverage factor k comes from Student's t for about 95 % (95.45 %) coverage; with large ν_eff, k = 2. U = k u_c.
Reading the budget
The "share" column shows each source's contribution to u_c². Reducing the largest contributors gives the most improvement.
Good practice
Avoid double counting (for example repeatability already included in a certificate value). Report the result as y ± U with k and the coverage probability.
A simple example of sensitivity
If a length L is measured at temperature t and corrected to 20 °C with expansion coefficient α, the sensitivity of the result to temperature is c = L × α. For a 100 mm steel part (α = 11.5 × 10⁻⁶ /K), c = 0.00115 mm/K.
Worked example
Six readings averaging 100.012 have a Type A uncertainty of 0.00058 with 5 degrees of freedom. With a reference standard of ±0.004 (k = 2, 50 dof), rectangular temperature and drift terms of ±0.003 and ±0.002, and 0.001 resolution, u_c = 0.00296, ν_eff ≈ 224, k = 2.01 and U = ±0.0060.
Limits
Linear models with uncorrelated inputs; correlated or strongly non-linear cases need GUM Supplement methods. Not an official accreditation document.
Exemple resolt
Six readings plus three Type B terms give U = ±0.0060 at k = 2.01.
En resum
- Categoria: Matemàtiques i ciència
- S'executa al navegador, sense registre, sense pujar res.
- La informació que introdueixes es queda al dispositiu.
- Disponible en anglès.
- Contingut revisat per última vegada:
- Font oficial: JCGM 100:2008 (GUM): standard uncertainties from Type A (s/√n) and Type B evaluations (rectangular a/√3, triangular a/√6, U-shaped a/√2, normal U/k), sensitivity coefficients, combined standard uncertainty u_c = √Σ(c_iu_i)², Welch-Satterthwaite effective degrees of freedom, coverage factor from Student t for 95.45 %, expanded uncertainty U = k·u_c; resolution uncertainty d/(2√3)
Preguntes freqüents
What is the GUM?
The international guide for evaluating and expressing measurement uncertainty.
What is the difference between Type A and Type B?
Type A uses statistics of repeated readings; Type B uses other information.
Why divide by √3?
That converts a rectangular half-width into a standard uncertainty.
When is k not 2?
When effective degrees of freedom are small; k comes from Student's t.
How should I report the result?
As y ± U with the coverage factor and coverage probability.
Is my data stored?