Measurement Uncertainty Budget Calculator (GUM)
Enter repeat readings, the display resolution and a table of Type B contributions with distributions, sensitivity coefficients and degrees of freedom to get combined and expanded uncertainty.
Εκτελείται στο πρόγραμμα περιήγησής σας. Τα αρχεία σας δεν φεύγουν ποτέ από τη συσκευή. Δείτε πώς να το ελέγξετε
…
Πώς γίνεται
- 1Paste repeat readings
For Type A.
- 2Enter resolution
Display resolution of the instrument.
- 3Fill the budget table
Value, distribution, divisor, sensitivity, dof.
- 4Choose the coverage factor
From t or fixed k = 2.
- 5Read the result
u_c, ν_eff, k and U.
Σχετικά με αυτό το εργαλείο
Every measurement result needs a statement of uncertainty, and the Guide to the Expression of Uncertainty in Measurement (GUM) is the international method for producing it. Calibration labs accredited to ISO/IEC 17025 present it as an uncertainty budget: a table of contributions, each converted to a standard uncertainty, combined and expanded with a coverage factor. This calculator builds such a budget, including Type A repeatability from your readings, the display resolution, the Welch-Satterthwaite effective degrees of freedom and the coverage factor.
Who this is for
Calibration and testing laboratory staff, quality engineers, metrology students and anyone preparing an ISO/IEC 17025 uncertainty statement.
Type A evaluation
From n repeat readings, the standard uncertainty of the mean is s/√n with n − 1 degrees of freedom. Paste your readings and it is added to the budget automatically.
Type B evaluation
Other sources come from certificates, specifications, handbooks or judgement. Convert each to a standard uncertainty: - **Normal**: expanded uncertainty from a certificate divided by its coverage factor (often k = 2). - **Rectangular**: half-width a divided by √3, for limits with equal probability, such as resolution or manufacturer specifications. - **Triangular**: a/√6, when values near the centre are more likely. - **U-shaped**: a/√2, for sinusoidal effects such as temperature cycling.
Resolution
A digital display with resolution d contributes d/(2√3). Enter the resolution and it becomes a rectangular term.
Sensitivity coefficients
Each input affects the result through a sensitivity coefficient c_i (a partial derivative or a unit conversion). The contribution is c_i u_i.
Combining and expanding
u_c = √Σ (c_i u_i)². The Welch-Satterthwaite formula gives effective degrees of freedom ν_eff = u_c⁴ / Σ((c_i u_i)⁴/ν_i), and the coverage factor k comes from Student's t for about 95 % (95.45 %) coverage; with large ν_eff, k = 2. U = k u_c.
Reading the budget
The "share" column shows each source's contribution to u_c². Reducing the largest contributors gives the most improvement.
Good practice
Avoid double counting (for example repeatability already included in a certificate value). Report the result as y ± U with k and the coverage probability.
A simple example of sensitivity
If a length L is measured at temperature t and corrected to 20 °C with expansion coefficient α, the sensitivity of the result to temperature is c = L × α. For a 100 mm steel part (α = 11.5 × 10⁻⁶ /K), c = 0.00115 mm/K.
Worked example
Six readings averaging 100.012 have a Type A uncertainty of 0.00058 with 5 degrees of freedom. With a reference standard of ±0.004 (k = 2, 50 dof), rectangular temperature and drift terms of ±0.003 and ±0.002, and 0.001 resolution, u_c = 0.00296, ν_eff ≈ 224, k = 2.01 and U = ±0.0060.
Limits
Linear models with uncorrelated inputs; correlated or strongly non-linear cases need GUM Supplement methods. Not an official accreditation document.
Λυμένο παράδειγμα
Six readings plus three Type B terms give U = ±0.0060 at k = 2.01.
Με μια ματιά
- Κατηγορία: Μαθηματικά & επιστήμες
- Εκτελείται στο πρόγραμμα περιήγησης, χωρίς εγγραφή, χωρίς μεταφόρτωση.
- Τα στοιχεία που εισάγετε μένουν στη συσκευή σας.
- Διαθέσιμο στα αγγλικά.
- Τελευταία αναθεώρηση περιεχομένου:
- Επίσημη πηγή: JCGM 100:2008 (GUM): standard uncertainties from Type A (s/√n) and Type B evaluations (rectangular a/√3, triangular a/√6, U-shaped a/√2, normal U/k), sensitivity coefficients, combined standard uncertainty u_c = √Σ(c_iu_i)², Welch-Satterthwaite effective degrees of freedom, coverage factor from Student t for 95.45 %, expanded uncertainty U = k·u_c; resolution uncertainty d/(2√3)
Συχνές ερωτήσεις
What is the GUM?
The international guide for evaluating and expressing measurement uncertainty.
What is the difference between Type A and Type B?
Type A uses statistics of repeated readings; Type B uses other information.
Why divide by √3?
That converts a rectangular half-width into a standard uncertainty.
When is k not 2?
When effective degrees of freedom are small; k comes from Student's t.
How should I report the result?
As y ± U with the coverage factor and coverage probability.
Is my data stored?
No. Everything is calculated in your browser.